In this vide we cover how to use a Fluke ESA 612 electrical safety analyzer. We demonstrate ground wire resistance testing, touch currents tests, and patient leakage tests.
The test performed are in line with NFPA 99 - AAMI/ANSI 99 testing standards
Ground wires resistance values should be less than 0.5 ohms
Touch current should be run in
No Fault Condition (ground intact) with the DUT in the on and off state less than 100 microamps
Single Fault Condition (ground lifted) with the DUT in the on and off state less than 500 microamps
Patient Leakage and Patient Auxiliary Leakage
No Fault Condition (ground intact) with the DUT in the on and off state less than 100 microamps
Single Fault Condition (ground lifted) with the DUT in the on and off state less than 500 microamps
Hope you enjoy the video make sure to subscribe to learn more!