There are maaany things missing from this but from the top of my head... - Wear suitable gloves to avoid touching the column with your bare hands. - When you cut the column with a wafer, you need to make a quick score with the smooth edge of the cutting wafer and essentially flick the end to reveal a clean cut. You should never need to bend the column or snap it. Always check the cut with a magnifier (jewellers loupe works well) to ensure the cut is clean. A poor cut will affect the chromatography. Wipe the end with a lint free cloth and solvent (methanol works) and always cut after inserting the ferrule to avoid a blockage. - With a new column, you need to purge at a low temp with a suitable carrier gas prior to conditioning to remove oxygen. This will vary based on the diameter of the column (narrower = longer purge time) You will also need to leak check your inlets prior to conditioning the column. Heating the column to a high temp with oxygen present will seriously reduce the column life or even destroy it before first use. These are some really basic principles of column insertion.
@joeleo9623 Жыл бұрын
Thanks for the info Is there more things to take care of ?
@davidcady54575 жыл бұрын
There are several things missing from this installation video, you should cut the column after installation of the ferrule, it is best practice to slide the ferrule on with the column end pointed down to prevent tiny ferrule fragments from traveling down the column past where you'll cut it. Then you must inspect your cut, it must be square - if not, cut it again. When cutting you just want to score the outer surface, then gently apply pressure to bend the capillary, it should break cleanly and with an even edge with very little pressure and bend angle, if you can still bend it, you have not scored it enough. If you choose to use no gloves it is best practice to always wipe the tip of the column with MeOH to remove contaminants, while you mention fingerprint oils, the truth is your hands are also covered with fragrances from hand soaps, oils from foods, and numerous other organics which yes may be baked off, but they will be baked off in your inlet and detector and cause you baseline headaches- best to just clean the end with MeOH. - This is especially true when using mass spectrometry instead of a simple FID detector. Your installation procedure also shows you using knurled finger tight nuts which are not standard and are not the ones you demonstrated on the bench. Those standard nuts should be fingertight then an additional 1/4 turn by a wrench to compress the ferrule, especially when using 85/15 ferrules and not 100% graphite. Finally, installation is not complete until you condition the column - it is my personal preference to always condition the column with only the inlet attached because I want to be as clean as possible. You must first purge the column/inlet of any air for about 10-15mins, then you can start heating the column at a nominal flow rate and ramping up to your conditioning temp and holding for several hours. - Then you can install the detector side and proceed with a Grob test injection to check proper installation and column activity.
@passengerman48055 жыл бұрын
U are right
@soddiq27085 жыл бұрын
Please make tutorial
@high5compliments454 жыл бұрын
Hello,i sell these things for very cheap. Please contact me if you need one
@davidcady54574 жыл бұрын
@@soddiq2708 I wish I could. I would love to have a dedicated channel for these types of tutorials, unfortunately, I no longer have access to a lab facility that would permit this. Maybe in the future...
@davidcady54574 жыл бұрын
@@high5compliments45 Are you referring to the 5890???
@MrBennyc1236 жыл бұрын
I would add that the column should get cut after the ferrule has been inserted onto the column, it is common for graphite to block the column. Very helpful video !
@tanveerkadri72155 жыл бұрын
I'm agree with you
@Edwin-md7dd5 жыл бұрын
@@tanveerkadri7215 We also do agree.... Very wrong!!!!! to cut before inserting into the ferrule.
@high5compliments454 жыл бұрын
I sell these things for very cheap. Please contact me if someone needs some.
@devprasad48046 жыл бұрын
Very very thank you for update this kind of concept.
@arulmuruganchemist26132 жыл бұрын
Thank you for teaching🥰india
@ozone70284 жыл бұрын
It's nice to see the ancient 5890 series X)
@michaelgrishko80283 жыл бұрын
It's robust, strong and friendly. Like Volkswagen Beetle from 70-s, it still can work, in contrast to modern fragile snowflake-tools, requiring repair each month
@jonheblanco27373 жыл бұрын
A cordial greeting, where I can obtain information on the frequency with which consumables or parts should be replaced either from the injection port, column or detector
@hichambenaissa31633 жыл бұрын
Thé cutting of the column must be done just before connecting the edges to injector and detector. In addition, it has to be wiped with soft cloth or a piece of cotton and methanol.
@shahnewajenayetkarim79923 жыл бұрын
Will the Septa remain after fixing the Nut or removed?
@farhanshaher30325 жыл бұрын
Thank you
@rajarvindsinh55434 жыл бұрын
Tnx madam🙏🏻
@dipakmochi4381 Жыл бұрын
Thak u
@sammawhamma6 жыл бұрын
This was SO HELPFUL thank you!!!!
@vashnator4 жыл бұрын
this video is SO wrong you're touching everything with your bare fingers leaving your bodily oils on the column and ferule, and plus cutting before you install nut and ferule... very bad practice, good luck with your baselines.
@high5compliments454 жыл бұрын
Hello,i have these things for sale. For less then what they actually cost. Maybe you need one?