If you get 4 beeps that means high pressure. what is needed to fix that problem? Also what if no air flows out and the tubing is new and no kinks what does that entail? Thanks
@Spygujjaryt8 ай бұрын
Did you get the solution I'm facing same problem
@somendrosingh59434 жыл бұрын
Machine is running for 4-5 min and off again automatically. What would be problem inside?
@Crazysummerdays3 жыл бұрын
I need some help, my machine keeps beeping every 10 mins.
@2ingrams2 жыл бұрын
Mine beeps 4 times in a row non stop. Did you figure out the issue
@catslivesmatter1268 Жыл бұрын
@@2ingramsmine drops after the beeping
@sharonm59232 жыл бұрын
I bought 2 of these 8 years ago....however there are 2 filters that need periodic cleaning....I did not know this until viewing another KZfaq video..the first set was coated with black powdery dust..they are drying now ...cleaning them should be addressed ..
@Rorschach00Testing4 жыл бұрын
Hello! My Devilbiss has 1700hrs on it, brand new. My noise meter says 80 db. That's way too high, no? Also, I removed the back cover and touched the motor. It was really hot, too hot to touch. Is this normal? Plus, what parts should I order ahead of time to avoid downtime? Thanks Calgary, Canada
@buckleyburgess45683 жыл бұрын
Does your machine need to be calibrated for altitude versus sea level? visionaire 5
@satishsiriyannavar47333 жыл бұрын
my vision air is not getting set for oxygen flow.its not increasing more then 1 ltr per minute.what is the cause.
@sandracasella73154 жыл бұрын
Donde puedo conseguir el motor de concentrador de oxígeno portátil airevision igual a ese en Montevideo Uruguay gracias bendiciones
@HuseyinAydin43526 жыл бұрын
kullanmaya bir aydır devam ediyorum cihaz güzel ama cihazın performansını bilmiyorum yüzde kaç oksijen veriyor henüz öğrenmedim.
@lajosorosz3 жыл бұрын
A kiserő szöveget magyar nyelven hol lehet meghallgatni .
@StuckInThe80ssss3 жыл бұрын
Ok so is this oxygen or no
@ahlama15748 жыл бұрын
I know it no need to put any o2 My machine it shows 0 letter per minute How can l fix it And what is the problem!??
@Oxygenguys5 жыл бұрын
First remove the tubing and/or humidifier bottle from the concentrator to make sure its not the tubing or humidifier restricting the flow.
@fuerzadelbien2.0familiafb775 жыл бұрын
porque se apaga a los 4 minutos
@sharadaindi27025 жыл бұрын
Will the alarm ring continuously until we fix the problem??
@Oxygenguys5 жыл бұрын
A solid alarm could mean, no power. Start by making sure there is power to the outlet and the machine is running. Remove all tubing and/or humidifier bottles. Make sure the desired flow is registering in the window. Also, good to have the oxygen content measured by a oxygen supplier to make sure the oxygen level is therapeutic.
@Oxygenguys5 жыл бұрын
Yes alarms will remain, until the problem is identified and solved.
@nazmiyesahin20435 жыл бұрын
Lütfen türkçe anlatın
@user-bn6xe4cl2d4 жыл бұрын
@@Oxygenguys كم السعر
@user-bn6xe4cl2d4 жыл бұрын
@@Oxygenguys كم السعر
@maziniqbal38464 жыл бұрын
It woukd have been nice if u had made us hear the beeps
@medicalequipmentspecialist65704 жыл бұрын
Will let the manufacturer aware of your request. Txs
@nerdyorganist9 жыл бұрын
Typically how many hours do you get from one of these units? I have the airsep elite 6lpm with dual flow meters. I've had this for about 12 years but I only use as needed so total hours are only 2500. However my oci indicates low oxygen at 5 lpm or above. I'm wondering is the sieve beds can need repoured just from age even without much usage.
@Oxygenguys9 жыл бұрын
+nerdyorganist Hello, sorry for the late response. The site is still under construction. I believe you are asking about how many hours does a unit typically last. I have seen thousands of hours on stationary oxygen concentrators. As long as the Oxygen content is greater than 90 percent, and the 02 flow is working. Some units have built in oxygen concentrator indicators that alarm the user when the oxgyen content drops below 85% more or less depending on the MFG. thanks for your comment
@Oxygenguys9 жыл бұрын
+nerdyorganist Yes, the higher the liter flow, the lower the 02 content. So I would have it analyzed at our presricribed liter flow. However, the OCI should turn off after 5 to 15 minutes at all liter flow. Have your DME or local biomedical company take a look.
@margitballa67765 ай бұрын
Szeretném ha magyarul hallhatnám youTube csatornán.
@2ingrams2 жыл бұрын
Mine beeps 4 times over and over again
@ahlama15748 жыл бұрын
Do I need to refill o2 How can I ?
@Oxygenguys5 жыл бұрын
You need someone to analzye the content of oxygen coming out of the concentrator to make sure its therapeutic.
@Verbatimorgish2 жыл бұрын
Our pos keeps beeping
@Oxygenguys2 жыл бұрын
Need a service center. Just because a unit runs doesn't mean it putting oxygen purity % >90% A home 02 analyzer prob would uncover the 02 % is low. Beeping and alarms are to protect patients, and that's a good thing. We hate them too, but they happen for a reason.
@Verbatimorgish2 жыл бұрын
@@Oxygenguys thank you. From what i read it may be compressor, which i may be able to free up obstruction from and fix. But if that fails, we're gonna have to do something. We are very poor.