In this episode Shahriar takes a look at a non-functional Fluke 5450A resistance calibrator box. The unit is equipped with a wide range of resistors ranging from 1-Ohm to 100-MOhm some of which are 0.005% tolerance. The instrument powers on and shows a single LED on the front panel without any other activity.
Upon further inspection it becomes clear that there are no other activities from the main CPU despite the clock signal being present. A replacement CPU returns the unit to normal front-panel operation. The analog board also has two blown fuses which may be the result of a in-rush event. After fuse replacement the analog relay arrays also function. The unit is measured against a Keysight 3458A 8.5-digit multimeter and shows remarkable accuracy despite its age.
www.TheSignalPath.com
/ thesignalpath​​​
www.Patreon.com/TheSignalPath

Пікірлер: 71
@tomking6006
@tomking6006 Жыл бұрын
One repairer I watch (I can't remember exactly who, but it may be Marco Reps) always takes out the eeproms and backs them up which seems like a very sound idea.
@Thesignalpath
@Thesignalpath Жыл бұрын
Yes, I should do that.
@vadimbellous8313
@vadimbellous8313 10 ай бұрын
Feedback loop does it
@silverXnoise
@silverXnoise Жыл бұрын
Love xDevs. Such a contribution to modern EE enthusiasts. Also, Ilia’s videos are like yours-a masterclass in top-notch engineering. Thanks to you both for sharing so much knowledge.
@HorochovPL
@HorochovPL Жыл бұрын
After last Kingpin lab tour, featured on Gamers Nexus channel, I tried to track down Illya "TiN" channel, with no results. Thanks for pointing me to xDevs!
@arthurmoore9488
@arthurmoore9488 Жыл бұрын
@@HorochovPL Similar story here. I was so glad to find he's doing well and all the interesting things he's been up to.
@EEVblog
@EEVblog Жыл бұрын
Nice! IIRC my unit was a bit out compared with my 6.5 digit meter at the time. Might have to reevaluate it.
@kuglepen64
@kuglepen64 Жыл бұрын
Some of the PPMs escaped. Don’t tell Marco Reps.
@OctavMandru
@OctavMandru Жыл бұрын
Shariar, it looks almost too easy. It's so refreshing to see this, it really makes us plaebeians think we could have it easy, as well. Thank you, repair videos are the best.
@JigilJigil
@JigilJigil Жыл бұрын
The guy has a Ph.D in Electrical Engineering, works at Bell Labs, with decades of experience, I assume it should be easy for him unlike us plebeians.
@amptechron
@amptechron Жыл бұрын
Always entertaining and informative. Thanks for a great channel!
@djjackson2200
@djjackson2200 Жыл бұрын
I love when your channel shows up on my notifications. Thanks for the awesome content!
@Edisson.
@Edisson. Жыл бұрын
Hi, great catch again, whenever I've bought older measuring equipment on Ebay it's been a service massacre except for the HP 427A and 209A - it suffered in shipping and the slot connector for the power supply board cracked. This catch belongs to the WOW category 😁, I would be quite interested in the process of damage to the circuit for the relay power supply. My thinking is that the cpu got damaged and sent incorrect data to control the relay and it overloaded the power circuit, I would definitely do a deep relay check. Thank you for your wonderful videos, you taught me a lot not only about repairs, but also about equipment that I did not know. Nice day 🙂 Tom
@dosgos
@dosgos Жыл бұрын
Mesmerizing relay action. I learned a lot in this short video!
@__--JY-Moe--__
@__--JY-Moe--__ Жыл бұрын
by digital ! the sexiest machines ever made! thanks 4 educating me, now I can fix the spaceshuttle! cheers!
@IanScottJohnston
@IanScottJohnston Жыл бұрын
Leave the 3458A powered 24/7 (30w approx.) if you aren't already, and perform an ACAL after a few days, then weeks, then months whilst doing some tests in between. It worked wonders for mine.
@randylemn
@randylemn Жыл бұрын
The missing chip is not used in the 5450A, The same CPU board was used in the 8520A and 8520A-PRT. Before you go and calibrate the unit as it has been sitting a while you should exercise the relays. Also when you do the calibration you will want to use the rear terminals.
@jaro6985
@jaro6985 Жыл бұрын
If he only uses the front terminals, wouldn't it make sense to calibrate to those?
@randylemn
@randylemn Жыл бұрын
@@jaro6985 The front terminals are Banana jacks while the back terminals are Copper Binding posts. The binding posts allow for a much lower EMF due to thermals where Banana plugs are great for normal connections but if he wants to make sub ppm measurements on the 5450A, The copper binding posts will make a much better connection.
@RyanUptonInnovator
@RyanUptonInnovator Жыл бұрын
Great video.
@buildstoys
@buildstoys Жыл бұрын
Love the repair videos!
@TomStorey96
@TomStorey96 Жыл бұрын
Regarding uncovered EPROMs, I've been doing a bit of an experiment over the past year. I left a couple of them out to see what effect ambient and direct sunlight would have on the contents. For the EPROM in direct sunlight (sitting on the window sill facing west here in the UK) it took approximately 1 month before I started noticing that some bits were erased. In the end it was about 3 months to fully erase back to all 1s. The other EPROM is sitting on my desk exposed only to ambient light. The desk is about a metre from the west facing window, so it still receives sunlight, but generally not directly - just whatever comes in through the window as well as lamps, my monitor, etc. Ambient. After what is now about 10 months, this EPROM is still verifying 100% against the initial contents that I dumped. I don't have a production quality programmer with which to do verification, but so far it doesn't seem like ambient light in a room is sufficient to cause any immediate harm, and I imagine some EPROMs tucked away inside a case with even less ambient light available are even less susceptible to erasure. Adrian's Digital Basement has also done an interesting experiment where he left an EPROM with the window covered only by a piece of Post-It note in an eraser overnight, and it also verified correctly the following day. So I kind of have a theory that the window doesn't need to be made 100% opaque, and that the top surface of an EPROM is simply the most convenient place to put a sticker, even if it is a simple paper sticker. I guess paper is sufficient to block any light with sufficient properties to cause erasure. I'll keep doing my experiment and see how long it takes for ambient light to have an effect on my EPROM, but it doesn't seem to be the "certain/imminent death" that some people make it out to be..?
@mikeselectricstuff
@mikeselectricstuff Жыл бұрын
Erasure needs shortwave UV, which is blocked by glass, so any indoor location is probably fine.
@TheDefpom
@TheDefpom Жыл бұрын
I did a video about this about 5 years ago, from memory it only took 1 week to corrupt the EPROM data having it sitting on a window sill… I am in NZ where UV is more of an issue.
@Thesignalpath
@Thesignalpath Жыл бұрын
Interesting. Thank you.
@TomStorey96
@TomStorey96 Жыл бұрын
@@mikeselectricstuff I'll have to try putting one outdoors and see what kind of difference that makes.
@anullhandle
@anullhandle Жыл бұрын
@@TheDefpom It's probably less of an issue now since task lighting isn't florescent. We always used the al sticky backed covers that you could indent with ver no.
@t1d100
@t1d100 Жыл бұрын
Nice!
@benedienst
@benedienst Жыл бұрын
We still have the 5450A in operation at Fluke. The dry current relais in it are a bit troublesome as they develop instability with time. Also the high ohm reed relais in "newer" units which are plastic encased become leaky.
@universeisundernoobligatio3283
@universeisundernoobligatio3283 Жыл бұрын
For relay contacts that develop high closed resistance which happens when they got old and don't switch much current. Get a say 1000uf low ESR cap charge it up to 12V, put it across the contacts and close the relay, there will be a spark, which will clean the contacts and drop its closed resistance. Good. For another 20 years.
@benedienst
@benedienst Жыл бұрын
@@universeisundernoobligatio3283 This will work on normal relays but when you do this on types which have a coating to be switched without a "wetting current" the issue can even become worse, especially when you look onto it from a point of contact repeatability which is the major concern here less the absolute contact resistance. And this is not wild guessing but actual experience as we tested this as a last resort measure before buying new specially manufactured relays.
@universeisundernoobligatio3283
@universeisundernoobligatio3283 Жыл бұрын
@@benedienst Interesting, my experience with this has been with relays switching into logic gates which draw no current therefore no spark. So for repeatable resistance the contact materials must not oxidize over years of use?
@benedienst
@benedienst Жыл бұрын
​@@universeisundernoobligatio3283 with logic gates you are working into devices which accept quite a large error margin in their levels. The relays of those calibrators are required to achieve contact repeatabilities in the single digit ppm range or below. Especially for the workstations we do the Fluke 57xx calibrators resistance verification by transfer method, the repeatability as to be better than 0.5ppm over many measurements and also perform good on drift over a 90 day period in a similar range. The further you go beyond the decimal sign the more impact even the tiniest imperfections get. This is also the reason why you often find more "aged" equiipment in primary laboratories instead of brand spanking new instruments. It's all in regards of predictability and good behavior. only then you are able to squeeze out the number of digits you need to achieve good uncertainties for highj end equipment like the Fluke 57xx Series calibrators.
@universeisundernoobligatio3283
@universeisundernoobligatio3283 Жыл бұрын
@@benedienst Thanks for the info, yes its can be a black art to get that type of precision and stability . Lots of experience plus lots of trial and error. I have many years experience doing strain gauge measurements for gas turbine R&D, designed and built many telemetry transmitters for rotating component strain gauge measurements.
@ivanr5315
@ivanr5315 Жыл бұрын
Very good video! Might be worth saving a copy of the eproms content just in case they ever become corrupted (assuming that they are socketed)
@the8bitpc_805
@the8bitpc_805 Жыл бұрын
Nice job. Actually, it's all Butch's doing, but since Butch can't talk, his master has to do the talking 🤫
@davidv1289
@davidv1289 Жыл бұрын
The large chip used for IEEE-488 support is a MC68488, a purpose built PIO chip designed by Motorola for the 6800 series microprocessors. It handles most of the bus handshaking and data transfer operations. Interestingly, it cannot drive the bus directly instead requiring separate bus transmitter/receiver chips. The MC3447 was the chip originally used for this purpose.
@rhirsch81
@rhirsch81 Жыл бұрын
6:20 interesting choice to use flex PCB material to connect that regulator to the PCB. They must have decided the labor to use individual wires would be more than the cost of that purpose built flex circuit bit.
@KallePihlajasaari
@KallePihlajasaari Жыл бұрын
The flex ribbon was probably a standard part from the first time it was used and became popular.
@hinz1
@hinz1 Жыл бұрын
Funny, one would expect such an ancient microprocessor with µm feature size to last forever. Same relays as in my Fluke 5440B, btw
@TheDefpom
@TheDefpom Жыл бұрын
I did a couple of repair videos on one of these, mine had issues with all of the relays, I spent ages taking them apart and cleaning them. You can tell if the relays are not perfect by switching values back and forth to see if the reading is exactly the same each time, I highly recommend you check for it.
@neonkev7866
@neonkev7866 Жыл бұрын
Awesome. Good thing old micros do not contain any firmware. What is the temp. coefficient on those nice resistors? I wonder if you can calibrate the old Fluke manually? I'm sure Fluke themselves would still cal it too, for a fortune of course.
@xDevscom_EE
@xDevscom_EE Жыл бұрын
Proper calibration of 5450A is rather tedious process, since it has very good uncertainties as calibrator, very similar to 1 year spec of calibrated 3458A. It's not as simple hook up calibrated DMM and record values, but whole setup with external sources, null-meters, etc. Today can be simplified a bit with bridges, but not much. Fluke goes into great details about it in 5450A manual too, which is excellent.
@neonkev7866
@neonkev7866 Жыл бұрын
@xDevscom_EE interesting info Illya. Sounds like a proper project for your lab full of calibrators and met gear! Fluke should do some lab and manufacturing video tours as marketing to engineers. Would be cool to see the who, where, and how behind world class metrology boxes. Not like it could be easily copied just by watching a video...
@olafzijnbuis
@olafzijnbuis Жыл бұрын
Interesting repair! You did not mention the type of processor you had to replace. Z80?
@KallePihlajasaari
@KallePihlajasaari Жыл бұрын
The pinouts on the schematic at 9:28 correspond to a Z80.
@olafzijnbuis
@olafzijnbuis Жыл бұрын
@@KallePihlajasaari Thanks. I am NOT surprised... It was a very common processor at the time. I had one in my TRS 80 Computer ca 1982
@seeigecannon
@seeigecannon Жыл бұрын
Are those going to be mercury wetted relays or something? I would think anything other than that would cause the resistance to go increase with time/use through the relay itself.
@mikeselectricstuff
@mikeselectricstuff Жыл бұрын
Looks like it's using a 4-terminal measurement, so relay resistance shouldn't be an issue, just need to ensure no thermocouple effects
@kirkpennock2997
@kirkpennock2997 Жыл бұрын
I wonder what keeps the relays contacts from getting crusty over time? Are they sealed with argon or neon and some kind of moisture absorber?
@Thesignalpath
@Thesignalpath Жыл бұрын
Yes, hermetically sealed.
@jimomertz
@jimomertz Жыл бұрын
How do they compensate for the contact resistance of the relay contacts?
@bradleyodens299
@bradleyodens299 Жыл бұрын
I think there is a 60Hz flicker on the display. I am not sure that is the way it was designed to perform, unless that only shows up in the video.
@jeanious2009
@jeanious2009 Жыл бұрын
Perhaps the contracts if the relays add a bit of resistance
@Thesignalpath
@Thesignalpath Жыл бұрын
In a 4-wire measurements the contacts are ignored.
@LeDabe
@LeDabe Жыл бұрын
The relative error seems to always be close to 1E-5 ~~
@simontay4851
@simontay4851 Жыл бұрын
9:08. What is the missing option that the unpopulated chip would be for? In other words, what does that chip do? Looks like another CPU by the size of it. Could you add it back?
@SidneyCritic
@SidneyCritic Жыл бұрын
It would've been good if we got a close-up of the micro, even though it's not going to be a Z80 - lol -. But then again, it's probably just some uncommon ancient micro that I'm not going to know anyway - lol -. At least we could've got a date code - lol -. A lot of lol's going on here - lol -.
@ctid107
@ctid107 Жыл бұрын
Looks like Motorola 6848
@KallePihlajasaari
@KallePihlajasaari Жыл бұрын
The schematic pinouts correspond to a Z80.
@ManofCulture
@ManofCulture Жыл бұрын
Calibrating calibrator using calibrator
@user-xr5el7tk6i
@user-xr5el7tk6i 8 ай бұрын
There is a schematic fluke 434?
@SergioCElias
@SergioCElias Жыл бұрын
Hello how are you? Do you have the fluke 435 service diagram?
@hpux735
@hpux735 Жыл бұрын
I think I have some of those resistors and the parts they're made of. A friend of mine's brother used to work at the Fluke factory in Washington. Let me know if you're interested and I'll mail them to you. I think you'll find them very interesting!
@meupaipapai7924
@meupaipapai7924 10 ай бұрын
Fluke 233 red ligth you now..
@iamgodlewski
@iamgodlewski Жыл бұрын
first
@TheOpticalFreak
@TheOpticalFreak 10 ай бұрын
Copper tape is the worst thing you can use! 🤦🏻‍♂️😝
@Thesignalpath
@Thesignalpath 10 ай бұрын
Why?
TNP #13 - Teardown & Repair of a Keithley 2001 7.5-Digit Multimeter
13:26
КАК ДУМАЕТЕ КТО ВЫЙГРАЕТ😂
00:29
МЯТНАЯ ФАНТА
Рет қаралды 10 МЛН
Can A Seed Grow In Your Nose? 🤔
00:33
Zack D. Films
Рет қаралды 27 МЛН
EEVblog #544 - Fluke 5450A Resistance Calibrator Teardown
36:51
RTO2014 Repair Part 1
23:31
jjoonathan
Рет қаралды 1,4 М.
The Most Important Circuit for our Electrical Future?! (PFC) EB#55
11:26
Lecture 1: Introduction to Power Electronics
43:22
MIT OpenCourseWare
Рет қаралды 322 М.
TSP #219 - Agilent 81101A 50MHz 20Vpp Pulse Generator Teardown & Repair
17:47
EEVblog #1038 - Uni-T UPO2104CS Oscilloscope Teardown
32:53
EEVblog
Рет қаралды 100 М.
Klavye İle Trafik Işığını Yönetmek #shorts
0:18
Osman Kabadayı
Рет қаралды 8 МЛН
КРУТОЙ ТЕЛЕФОН
0:16
KINO KAIF
Рет қаралды 6 МЛН
8 Товаров с Алиэкспресс, о которых ты мог и не знать!
49:47
РасПаковка ДваПаковка
Рет қаралды 175 М.
Мой новый мега монитор!🤯
1:00
Корнеич
Рет қаралды 199 М.
Сколько реально стоит ПК Величайшего?
0:37