Пікірлер
@gustavoferrari1913
@gustavoferrari1913 13 күн бұрын
Preciso restar o titulador para ficar com a configuração de fábrica
@gustavoferrari1913
@gustavoferrari1913 13 күн бұрын
Tenho uma dúvida qual a senha para acessar a função serviço?
@MetrohmTV
@MetrohmTV 8 күн бұрын
Entre em contato com sua subsidiária local ou distribuidor exclusivo da Metrohm para obter suporte técnico.
@BenoitPlomb-u8l
@BenoitPlomb-u8l 17 күн бұрын
What maintenance intervals (calibration, electrolyte replacement) would you recommend for a pH electrode that is used very rarely, e.g. once a year?
@MetrohmTV
@MetrohmTV 8 күн бұрын
Hi This very much depends on the accuracy of the result you want to achieve. For example, if the result is only a rough guide, it is fine to perform a calibration or replace the electrolyte before the pH measurement. However, if you want the result to be as accurate as possible, maintenance is more important and should be carried out at least monthly or even weekly. In general, frequent maintenance of the electrode is recommended for warranty reasons and also with regard to the lifetime of an electrode.
@BenoitPlomb-u8l
@BenoitPlomb-u8l 8 күн бұрын
@@MetrohmTV Thanks a lot for your answer !
@MetrohmTV
@MetrohmTV 8 күн бұрын
@@BenoitPlomb-u8l You're welcome. We hope this helps!
@rahnpincas142
@rahnpincas142 20 күн бұрын
Hi, our mining lab has just purchased an Eco Titrator 10080010 to be used for copper analysis via iodimetry. Can someone share a reference on how to set up the method using Eco Titrator? we have been using manual buret for iodometry, so this one is basically an upgrade. Thank you!
@MetrohmTV
@MetrohmTV 8 күн бұрын
Hi Your local Metrohm subsidiary or exclusive distributor can help you with this! Have you been in touch with them?
@ZogeriaSystems
@ZogeriaSystems 21 күн бұрын
On 913 with flexible sleeve how much electrode do we fill ? Thank you in advance
@MetrohmTV
@MetrohmTV 8 күн бұрын
Hi If you're talking about electrolytes, the electrode should always be filled with electrolyte up to the filling opening, regardless of the type of Metrohm pH electrode. You can find more information on this topic in the monograph 81085081EN "pH Measurement" under "Electrolyte" and "Electrolyte Outflow": www.metrohm.com/en/products/8/1085/81085081.html
@mohamedzaytoun168
@mohamedzaytoun168 29 күн бұрын
What if the blank read value of chloride.. Possible reason
@MetrohmTV
@MetrohmTV 29 күн бұрын
We assume that you see a Cl peak in the blank. You could try increasing the rinsing volume. Or run several blanks in a row to see if the peak decreases. If yes, then increase the rinsing volume. If not, then it could be an impurity in the chemicals or the ultrapure water. Whenever possible, you should use chemicals and ultrapure water without impurities. Please also check our Product Help Center IC: www.metrohm.com/en/service/product-help-center/ion-chromatography.html
@mohamedzaytoun168
@mohamedzaytoun168 29 күн бұрын
When i need to exchange
@MetrohmTV
@MetrohmTV 29 күн бұрын
The tube needs to be replaced when it's contaminated or defective.
@prakashsekarspk3716
@prakashsekarspk3716 Ай бұрын
How much price
@MetrohmTV
@MetrohmTV 29 күн бұрын
Hi Thank you for your comment. For instrument quotes, please contact your local Metrohm subsidiary or exclusive distributor: www.metrohm.com/en.html#country-selection Best regards Michel, Metrohm International Headquarters
@SimonLiangForever
@SimonLiangForever Ай бұрын
Subject: Price and Delivery Inquiry for Metrohm 18000010 (Quantity: 1) Dear Metrohm Sales Team, I am writing to you today on behalf of Omron Electronics GmbH, located in Langenfeld, Germany. We are interested in obtaining a price quote and estimated delivery time for one unit of the Metrohm 18000010 drive with write/read hardware for intelligent Dosing Units. We are currently in the planning phase for a new chemical analysis laboratory and are considering your product for precise reagent addition in titrations, pH adjustments, and sample preparation. We believe the Metrohm 18000010 could be a valuable asset in ensuring the accuracy and efficiency of our analytical processes. In addition to the price and delivery information, we would also appreciate details regarding your payment terms. As the end user of this product, we are committed to making a well-informed decision based on the comprehensive information you provide. Thank you for your time and attention to this matter. We look forward to your prompt response and the opportunity to discuss this further. Sincerely, David Remo | Purchase Manager Omron ELECTRONICS GmbH E: [email protected] Tel: +49 1744257779 A: Elisabeth-Selbert-Strasse 17, 40764 Langenfeld, Germany
@Mr239shiva
@Mr239shiva Ай бұрын
Dear Team, please provide contact info
@MetrohmTV
@MetrohmTV Ай бұрын
Hi You can contact your local Metrohm organization or exclusive distributor through our website: www.metrohm.com/en.html#country-selection Best regards Michel, Metrohm International Headquarters
@tnktxn640
@tnktxn640 Ай бұрын
Why do I have to calibrate more often when performed at high temperatures? If I do it at 28 degrees Celsius, is calibration necessary every day?
@MetrohmTV
@MetrohmTV Ай бұрын
We would recommend calibrating the sensor daily and at the temperature of the samples for exact results. It depends a bit on the accuracy that is required.
@arseniotedra4573
@arseniotedra4573 Ай бұрын
www.youtube.com/@arseniotedra4573 #ibelieve thankyou so much ❤️
@drmoatazali1297
@drmoatazali1297 Ай бұрын
Price please
@MetrohmTV
@MetrohmTV Ай бұрын
Hi Thank you for your comment. If you would like to receive a quote for this instrument, please contact your local Metrohm subsidiary or exclusive distributor. You can contact them through our website: www.metrohm.com/en.html#country-selection BR Michel, Metrohm International Headquarters
@nanakhvedelidze3745
@nanakhvedelidze3745 2 ай бұрын
which buffer i have to use for nitrate measurment?
@MetrohmTV
@MetrohmTV 2 ай бұрын
Hi You can find our recommendation in the electrode manual on page 8: www.metrohm.com/content/dam/metrohm/shared/documents/fact-sheets/81098042EN.pdf BR Michel, Metrohm International Headquarters
@John_K2
@John_K2 2 ай бұрын
I'm a great admirer of 884 professional VA.
@MetrohmTV
@MetrohmTV 2 ай бұрын
That's great to hear, thank you! We're happy you like the instrument!
@manojmodi414
@manojmodi414 2 ай бұрын
Thank you very much Sir
@zzzluvsk
@zzzluvsk 2 ай бұрын
Thanks. Always wondered how these work
@MetrohmTV
@MetrohmTV 2 ай бұрын
Great! We're happy you found this useful!
@Radhe_Tyagi
@Radhe_Tyagi 2 ай бұрын
Nice video...
@MetrohmTV
@MetrohmTV 2 ай бұрын
Thanks!
@hiteshkale6044
@hiteshkale6044 3 ай бұрын
Operate your system
@hiteshkale6044
@hiteshkale6044 3 ай бұрын
Please provide how to ion chromatography of your system.
@MetrohmTV
@MetrohmTV 3 ай бұрын
Hi Thank you for reaching out to us. In order to get accurate and comprehensive information on how to operate our IC systems, we recommend that you contact your local Metrohm subsidiary or exclusive distributor for support. Alternatively, you can also find information in the manual of your IC system. BR Michel, Metrohm International Headquarters
@wagstercarlos1525
@wagstercarlos1525 3 ай бұрын
Hello, does this cleaning prejudice the calibration?
@MetrohmTV
@MetrohmTV 3 ай бұрын
Hello, this does not affect the calibration. After cleaning, you need to prep the burette with fresh titrant and your system will be working correctly.
@RamsilPrivateLimited
@RamsilPrivateLimited 3 ай бұрын
Contact Details , We are Located Bangalore Karnataka
@MetrohmTV
@MetrohmTV 3 ай бұрын
Hi Please find the contact details of Metrohm India here: www.metrohm.com/en_in/company/how-to-find-us.html
@ericksanchezperez829
@ericksanchezperez829 4 ай бұрын
What is the problem when the slope is Less than 97%? And pH(0) is higger than 7.10?
@MetrohmTV
@MetrohmTV 4 ай бұрын
Hi Erick There could be various reasons for this. You can find additional information on this topic in our Application Bulletin AB-188: www.metrohm.com/en/applications/ab-application-bulletins/ab-188.html If you need to troubleshoot your instrument or sensor, please get in touch with your local Metrohm subsidiary or exclusive distributor: www.metrohm.com/en.html#country-selection BR Michel, Metrohm International Headquarters
@akakisimonishvili7471
@akakisimonishvili7471 4 ай бұрын
Hello, I have a new HACH PHC101 electrode, but I can not calibrate, it says (standard not recognized. Rinse the electrode and repeat). Can you help me and tell me the reason?
@MetrohmTV
@MetrohmTV 4 ай бұрын
Hi Thank you for your comment. Could you please check if another electrode can be calibrated? This way we would know if the problem lies with the instrument or the sensor. In general, if you require instrument support, please contact your local Metrohm organization or exclusive distributor. BR Michel, Metrohm International Headquarters
@ramamohankilli5548
@ramamohankilli5548 4 ай бұрын
On what cases Water content result came zero, but the regular sample trend is 15%
@MetrohmTV
@MetrohmTV 4 ай бұрын
Hi Thank you for your comment. This problem could be due to a variety of reasons. Therefore, to get to the bottom of this, we recommend that you contact your local Metrohm organization or exclusive distributor. BR Michel, Metrohm International Headquarters
@manoranjansahoo8586
@manoranjansahoo8586 5 ай бұрын
Hello,pl suggest my tbn result is wrong
@MetrohmTV
@MetrohmTV 5 ай бұрын
Hi Please get in touch with your local Metrohm subsidiary or exclusive distributor. They will gladly help you troubleshoot these issues: www.metrohm.com/en.html#country-selection BR Michel, Metrohm International Headquarters
@hamidrezavaezi4431
@hamidrezavaezi4431 5 ай бұрын
Metrohm 827 ph lab cal
@MetrohmTV
@MetrohmTV 5 ай бұрын
Hi If you're looking for more information on electrode calibration, check out our Electrode Help Center: www.metrohm.com/en/service/product-help-center/electrodes/maintenance-and-care-of-electrodes.html If you're looking for instructions on the 827 pH lab, you can find the manual here: www.metrohm.com/content/dam/metrohm/shared/documents/manuals/88/88278001EN.pdf And if you need more support with your instrument, please contact your local Metrohm organization or exclusive distributor: www.metrohm.com/en.html#country-selection Best regards Michel, Metrohm International Headquarters
@kersanim1519
@kersanim1519 5 ай бұрын
Thank you
@MetrohmTV
@MetrohmTV 5 ай бұрын
You're welcome!
@KARIMA-xl3gj
@KARIMA-xl3gj 5 ай бұрын
I need to analyze lead photometrically using the optrode. I find it difficult to configure the method on the Omnis software. Any help ?
@MetrohmTV
@MetrohmTV 5 ай бұрын
Hi! Thank you for your comment. Your local Metrohm subsidiary or exclusive distributor can help you to configure your method. You can find the contact details on our website: www.metrohm.com/en.html#country-selection
@refar1624
@refar1624 5 ай бұрын
Muy buen video y lo explica muy bien.
@vinhtruongtv5727
@vinhtruongtv5727 6 ай бұрын
Chú mình mua thuốc không nhãn mác , ko có hạn sử dụng , mình muốn hỏi mình có thể giám định thành phần thuốc này ở đâu ? Và chi phí giám định này cao ko ?
@MetrohmTV
@MetrohmTV 5 ай бұрын
Một loại thuốc không nhãn mác, không hạn sử dụng sẽ không được lưu hành và khuyên cáo người sử dụng không nên dùng. Xác định thành phần thuốc rất khó khăn, phải dùng đến các loại máy phân tích hiện đại độ chính xác cao và giá thành rất đắt. Vì thế sẽ rất ít các cơ quan nhận mẫu kiểm nghiệm trong trường hợp này!
@prashantpatil5766
@prashantpatil5766 6 ай бұрын
Fantastic system
@bhagyaranasingha2216
@bhagyaranasingha2216 6 ай бұрын
Nice explanation 👌
@MetrohmTV
@MetrohmTV 6 ай бұрын
Thank you 🙂
@fatinrun
@fatinrun 6 ай бұрын
How to fix blockage piston ? Tq im highly appreciate if u can reply
@MetrohmTV
@MetrohmTV 6 ай бұрын
Hi @fatinrun A blocked piston on an exchange unit is usually caused by a blocked tube or anti diffusion tip. We suggest: - Remove the anti diff. tip and try dosing again - If the anti diff. tip doesn’t rectify the problem then remove the burette tip and dose again - If the blockage still occurs then disconnect the outlet tubing from the valve unit and try dosing - make sure you have tissue at hand to catch the dispensed liquid - If problem remains then the issue has to be caused by a blocked valve/stop cock which will need removing/rinsing and potentially replacing - If this doesn't fix the issue, please send an email to [email protected] and we will be happy to assist further.
@FormWerkstoff
@FormWerkstoff 6 ай бұрын
Ich weiß nicht genau warum meine Augendrüsenzellen leicht Flüssigkeit hergestellt haben ich werde aber darüber nachdenken. Bis heute kannte ich dieses Unternehmen nicht. Die LED Licht und Millibar Elektromebrane wurden von meinen Gehirn Psychologisch und Elektrochemisch verarbeitet.
@boobalanarjun2133
@boobalanarjun2133 6 ай бұрын
Why is there moisture parameter in most of the analysis? How the quality is analyzed with moisture parameter alone?
@MetrohmTV
@MetrohmTV 6 ай бұрын
Hi Moisture content is definitely one important parameter for the quality assurance, but it is not the only one. Next to that, multiple key parameters, like ethanol, glycerol, or active ingredients like sodium laureth sulfate can be measured simultaneously in less then one minute using the DS2500 Solid Analyzer. The Vision Air software displays all parameters on one screen when the analysis is finished.
@IsabelVasquez-ey5zx
@IsabelVasquez-ey5zx 6 ай бұрын
Is it water they use to clean?
@MetrohmTV
@MetrohmTV 6 ай бұрын
You should use distilled or ultrapure water for cleaning.
@samermajeed9975
@samermajeed9975 7 ай бұрын
Is ASTM D1533 available with the same injection method?
@MetrohmTV
@MetrohmTV 6 ай бұрын
Hi Thank you for your question. No, the oven method is not mentioned in ASTM D1533. It describes direct coulometric Karl Fischer titration. The samples must be added directly into the titration cell.
@ymjdea
@ymjdea 7 ай бұрын
@Nur_Md._Mohiuddin_Chy._Toha
@Nur_Md._Mohiuddin_Chy._Toha 7 ай бұрын
👍👍👍👍👍
@CJP3
@CJP3 7 ай бұрын
Thanks for the video!
@MetrohmTV
@MetrohmTV 7 ай бұрын
You're welcome!
@CJP3
@CJP3 7 ай бұрын
@@MetrohmTV I’ve been publishing on the power of field-able chemical analysis. I’ve been really impressed by what you’ve accomplished with your portable NearIR source.
@MetrohmTV
@MetrohmTV 7 ай бұрын
Thank you, we appreciate it!🙏
@EzraOyelakin
@EzraOyelakin 7 ай бұрын
Is the resin doing the exchange
@MetrohmTV
@MetrohmTV 7 ай бұрын
Yes. More precisely, the charged ion exchange groups on the surface of the resin are responsible for the ion exchange.
@quimicafeliz1784
@quimicafeliz1784 8 ай бұрын
Excelente
@study-pulse
@study-pulse 8 ай бұрын
Waow. What a presentation. Very clear and extraordinary. On point and understandable
@MetrohmTV
@MetrohmTV 7 ай бұрын
Thanks! We're happy you like it. BTW: You can find more recorded webinars in our Webinar Center: www.metrohm.com/en/service/webinar-center.html
@r.a.3679
@r.a.3679 8 ай бұрын
Brazil
@MetrohmTV
@MetrohmTV 8 ай бұрын
Hi If you're interested in our products and services, please contact our colleagues at Metrohm Brazil. www.metrohm.com/pt_br.html
@samueltinku250
@samueltinku250 10 ай бұрын
Great video 😊! Why only 4,7,9 buffers used for calibration! What is the maximum limit can dip the ph electrode or need to only dip the end of the electrode
@MetrohmTV
@MetrohmTV 10 ай бұрын
Hi Samuel Thank you for your comment. You always have to calibrate with buffers which lie within your measuring range. With buffers pH 4, 7, and 9, you’re already covering quite a broad pH range. Should you need to cover an even broader range, you can use pH buffers from other manufacturers. As for the electrode: The electrode has to be dipped in in such a way that the diaphragm is fully covered by your measuring solution. I.e., the glass membrane and diaphragm have to be immersed but the electrolyte filling hole has to stay above the sample. Best regards Michel, Metrohm International Headquarters
@samueltinku250
@samueltinku250 10 ай бұрын
@@MetrohmTV 😀 thanks for your reply 😊
@pastelpinks
@pastelpinks 10 ай бұрын
Hi what if there is backflow of cell solution back into the titrant Comp T5 tube and instead of dosing titrant it is dosing the cell solution back into the cell
@MetrohmTV
@MetrohmTV 10 ай бұрын
Hi We cannot tell what the problem might be without further information. Please contact your local Metrohm subsidiary or exclusive distributor for technical support in this matter: www.metrohm.com/en.html#country-selection Best regards Michel, Metrohm International Headquarters
@aswanthp9933
@aswanthp9933 10 ай бұрын
Can i use optrode for turbidity titration(CTAB surface area method)
@MetrohmTV
@MetrohmTV 10 ай бұрын
Hi We do have an application where we used the Optrode for turbidity titrations. However, turbidity titration does not work in every case. It depends on the difference of turbidity and on your sample. Therefore you would need to test it out for your specific application. Please find the application here: www.metrohm.com/en/applications/application-notes/aa-t-001-100/an-t-055.html If you need more support, feel free to contact your local Metrohm subsidiary or exclusive distributor: www.metrohm.com/en.html#country-selection Best regards Michel, Metrohm International Headquarters
@aswanthp9933
@aswanthp9933 10 ай бұрын
Thank u for your kind rply. But in our case the end point is not capturing at the equivalnace point.(we are using optrode equiped with 905 titrando)
@MetrohmTV
@MetrohmTV 10 ай бұрын
As mentioned in the last comment, the detection of the turbidity is dependent on your sample. Please get in touch with your local Metrohm Agency to resolve the issue: www.metrohm.com/en.html#country-selection Best regards Michel, Metrohm International Headquarters