Thanks for your video. What should I do if my Elcometer 319 and 116C show 2 significant different results?
@ElcometerOfficialChannel3 жыл бұрын
Thank you for your comment. As the Elcometer 319 and Elcometer 116C work in different ways, and the Elcometer 319 has an accuracy statement of ±0.5ºC (±1ºF), it is not uncommon that they produce different results. Also, as you are whirling the Elcometer 116C, while the Elcometer 319 remains static and can be placed on the surface to be painted, typically you will not be measuring the same pocket of air. For more information please contact techsales@elcometer.com or visit www.elcometer.com/contact-us to find your local Elcometer distributor.
@thicongson43828 ай бұрын
Please tell me which standard regulates the delta T index > 3 0C
@ElcometerOfficialChannel8 ай бұрын
Thank you for your comment, The standard you are referring to would be the ISO 8502-4 standard. Unless otherwise agreed, the steel surface temperature should generally be at least 3 °C above the dewpoint when paints are used. If you have any other enquiries, please email support@elcometer.com and they will do their best to help.
@Dingzerovn3 жыл бұрын
Currently, my elcometer 319 cannot start, the screen is black with a symbol of lightning and slash. So what could the error be?
@ElcometerOfficialChannel3 жыл бұрын
Thank you for your comment, and we are sorry to hear about your Elcometer 319. It may just require a software patch that we can send over to you. To help resolve your issue, please email techsales@elcometer.com and they will do their best to help you.
@Dingzerovn3 жыл бұрын
@@ElcometerOfficialChannel Thanks a lot for your support
@justryn19957 ай бұрын
The delta T needs to be 5° higher than what?
@ElcometerOfficialChannel7 ай бұрын
Hi, The Delta T needs to be greater than 3 degrees Celsius or 5 degrees Fahrenheit. If you have any more questions please get in touch at support@elcometer.com