Hello everyone, in this video, I will demonstrate LCMS triple quadrupole system maintenance, mainly for Shimadzu LCMS-80xx series models. From the video, you will see how simple and easy to do a regular LCMS maintenance, without vacuum shutdown. After turning off MS, waiting for the temperature cool down, then we can start the hands-on.
There are three sessions in this video, including interface cleaning, capillary replacement and how to change the DL without vacuum shut down.
Interface cleaning, first, open the unit and disconnect the PEEK tubing. Then slide the knob to remove the whole unit. Be careful to uninstall it after the temperature has been cool down.
Put the ion source unit carefully on the table. And prepare LCMS grade methanol and Kimwipe tissue for ion source cleaning. Wet the tissue with methanol and wipe off the soiling and any contamination on the surface of ion source. Please remember, do not leave any tissue residues on the surface. After that, we can continue to clean the sampling cone on the heated block. Loosen the screws using the Allen key provided. Loosen one screw first, then change to another one. Insert the drawing tool provided under the heated block, remove the heater flange out slowly. Using a screwdriver to loosen the screws on the sampling cone, then put this sampling cone into a clean beaker with methanol inside for sonication.
ESI interface capillary replacement. To replace the capillary, first disconnect the PEEK tubing. Then remove the knurled screw. Loosen and remove the ESI coupling, and remove the capillary ASSY. Take it gently to avoid bending it accidently. Hold the ESI pipe Assy carefully, make the alignment of ESI pipe with the nut. If it is confirmed clogged, change it with a new one. After that, mount the ESI capillary ASSY. Put the ESI coupling back by dropping it gently. Do not push it hardly. Remember to adjust the projection of the tip of capillary to be about 0.5-1mm. Tighten the knurled screw by hand. Re-connect the PEEK tubing.
Next is the ion source installation. Put the ion source unit back by insert it into the two bolts.
Last session is showing you how to change a DL. DL can be changed when it is contaminated. To change a DL, first disconnect the connector. Then using the Allen key to loosen the locking screws. After that, turn the notched section to the direction where the DL is to be removed, then draw it out. This is the DL. DL can be cleaned internally and externally using methanol. After cleaning, mount the DL back to the instrument. Turn the notched section in the direction for securing the DL to the heater flange. Then tighten the two DL locking screws using the Allen key provided. Please tighten the locking screws sufficiently. The last step is to re-connect the connector.
#Shimadzu_MS
#Triple_Quadrupole_LCMS
#MS_maintenance

Пікірлер: 2
@chebliyassir803
@chebliyassir803 2 ай бұрын
Thank you so much for the demonstaration ! It was very Helpful
@uazhar2013
@uazhar2013 Ай бұрын
How do you clean the quadruples
How to set up MRM method for veterinary drug compounds on LC/MS/MS
6:49
Shimadzu Asia Pacific
Рет қаралды 12 М.
Quickly understand various types of LCMS
7:59
Shimadzu Asia Pacific
Рет қаралды 9 М.
لااا! هذه البرتقالة مزعجة جدًا #قصير
00:15
One More Arabic
Рет қаралды 51 МЛН
Meet the one boy from the Ronaldo edit in India
00:30
Younes Zarou
Рет қаралды 16 МЛН
Stay on your way 🛤️✨
00:34
A4
Рет қаралды 33 МЛН
Kind Waiter's Gesture to Homeless Boy #shorts
00:32
I migliori trucchetti di Fabiosa
Рет қаралды 12 МЛН
How to Clean the Ionization Source on a Shimadzu Triple Quadrupole Mass Spectrometer
5:18
Shimadzu Scientific Instruments
Рет қаралды 920
How it works  - Agilent 6495C triple quadrupole LC/MS
3:40
Agilent Technologies
Рет қаралды 806 М.
LC-MS/MS Education Series: Quadrupole Theory and Use
6:51
Waters Corporation
Рет қаралды 50 М.
MassHunter GC/MS MRM optimizer
12:39
Agilent Technologies
Рет қаралды 3,2 М.
Installing a New GC Column into Your Inlet
2:59
Agilent Technologies
Рет қаралды 9 М.
Tandem Mass Spectrometry
13:26
ShortChemistry
Рет қаралды 46 М.
لااا! هذه البرتقالة مزعجة جدًا #قصير
00:15
One More Arabic
Рет қаралды 51 МЛН